SDD Ref. Haier HTC-40SDD E003/075

Solar Direct Drive Refrigerator for Storage of Vaccines

Haier HTC-40SDD E003/075

Ambient operating temperature: Min. +5°C, Max. +43°C

Net vaccine storage volume: 22.5 L

Supplied as a complete plug & play solar system including a temperature
monitoring device.

Incoterms

CIP

$2,400.00

Technical Guidance:

Contact UNICEF Supply Division, HTC, Cold Chain Unit for technical
questions. CPHHQ-SD-Cold Chain Unit < coldchainunit@unicef.org >

General Description:

Solar Direct Drive Refrigerator for Storage of Vaccines

Model: Haier HTC-40SDD

PQS ref. E003/075

Technical specifications:

A complete stand-alone SDD refrigerator and all necessary components and
fittings (Plug and Play).

Refrigerator cabinet:

Net vaccine storage volume: 22.5 liters

Refrigerant type: R600a

Ambient operating temperature: Min. +5°C, Max. +43°C

Freeze Protection: Grade A

Solar Power System:

Shall match the refrigerator power consumption, climate, ambient
temperatures and solar radiation reference source of 3.5KWh/m2/24 hours.

Country of origin for solar panels; China.

Supplied with:

1.Temperature Monitoring Device - 30 day electronic refrigerator
temperature logger:

Haier HETL-01, E006/042 attached to internal refrigerator basket

OR

Upon request: Either Berlinger Fridge-Tag 2E, E006/040 with an internal
sensor attached to internal refrigerator basket or Fridge-Tag 2E with
external sensor, E006/040 shall be supplied.

2.ON/OFF power switch, accessible to the user, is provided on the
appliance that isolates refrigerator from the solar power system.

3.10 fuses of each type are included.

4.Lid with lock, plus 2 keys and 2 denser wire baskets.

Manuals:

User, Installation and Technician’s manual supplied in English.

Shipping Information:

Packaging and labelling;

SDD refrigerator package: carton package with plywood pallet

Solar panel package: plywood

Estimated shipping weight: 155 kg (combined weight refrigerator & solar
array)

Estimated shipping volume: 1.20 m³ (combined volume refrigerator & solar
array)

Warranty:

1.The product is covered by a two-year replacement warranty in the
event of any component failure arising from defective design, materials
or workmanship.

2.Solar module power output: Warranty of 25 years to 80% of initial
power rating.

NOTES:

1.In case of any import restrictions regarding country of manufacture,
CO should contact SD and confirm solar panel, country of origin.

2.French & Spanish manuals are available on request.

3.A label, displaying vaccine storage advice & user instructions with
pictures is permanently mounted on cabinet.

4.One set of spare parts is recommended but not compulsory for every
10 units of SDDs.

5.A Basic Installation Toolkit is available on request. (For contents
of the toolkit, see supply web catalogue.)

6.Maintenance tool kits are available on request. (For contents of the
toolkit, see supply web catalogue.)

7.Further information for comparable products, can be obtained from
CPHHQ-SD-Cold Chain Unit < coldchainunit@unicef.org >.

Useful links:

1.Cold Chain
Support Package:

2. < a href="https://supply.unicef.org/unicef_b2c/app/displayApp/(layout=7.0-12_1_66_67_115&carea=%24ROOT)/.do?rf=y">>Supply
Catalogue:

Additional information

Incoterms

CIP

Reviews

There are no reviews yet.

Only logged in customers who have purchased this product may leave a review.

Vendor Information

  • Store Name: UNICEF Supply Division
  • Vendor: Claudia Lago
  • Address: Oceanvej 10-12
    2150 Nordhavn
    Denmark
  • No ratings found yet!

Product Enquiry